Charles Explorer logo
🇬🇧

Determination of nitrophenols by flow injection analysis with amperometric detection on boron doped diamond dilm electrode

Publication at Faculty of Science |
2010

Abstract

Boron doped diamond film electrode (BDDFE) was employed in amperometric detection cell in “wall jet” arrangement for determination of 2-nitrophenol (2NP), 4-nitrophenol (4NP) and 2,4-dinitrophenol (DNP) in flow injection analysis. Under optimized conditions (run buffer 0.01 mol.L−1 sodium hydroxide; flow rate 4 mL.min−1; detection potential 1 V; injection volume 20 μL), the limits of quantitation of 2×10−6 mol.L−1 were achieved.

The repeatability of the detector response is satisfactory (relative standard deviation around 2 % for c(2NP, 4NP and DNP) = 1×10−4 mol L−1) even without any BDDFE treatment