Practical method transfer from high performance liquid chromatography to ultra-high performance liquid chromatography: The importance of frictional heating

Publication at Faculty of Pharmacy in Hradec Králové |
2011

Abstract

The amount of frictional heating in UHPLC was empirically estimated. Frictional heating was measured for 1.0 and 2.1 mm I.D.

UHPLC columns. The heating was equal to +4 degrees C, +8 degrees C and +16 degrees C at 300, 600 and 1000 bar, respectively.

The frictional heating effect is still important with 1.0 mm I.D. columns. The best solution to compensate frictional heating is to decrease temperature in UHPLC.