The retention behavior of several series of free alpha- and omega-amino acids and positional isomers of amino pentanoic acid in the hydrophilic interaction chromatography mode (HILIC) was studied. The study was carried out on three stationary phases followed by post-column derivatization with fluorescence detection in order to describe the retention mechanism of the tested amino acids.
The effect of chromatographic conditions including acetonitrile content in the mobile phase, mobile phase pH (ranging from 3.5 to 6.5) and concentration of buffer in the mobile phase was investigated. The effect of the number of carbon atoms (n(C)) in aliphatic chains of the individual homologue of alpha- and omega-amino acids and the logarithm of the partition coefficient (logD) on retention was also a part of the presented study.
A good correlation (r > 0.98) between the logk and logD values of amino acids or n(C), respectively, was observed. The described linear relationships were subsequently applied to predict the retention behavior of individual members of the homologous series of amino acids and to optimize the mobile phase composition in HILIC.
The obtained results confirmed that the retention mechanism of alpha-amino acids, omega-amino acids and positional isomers of amino acids was based on the logD values and the number of carbon atoms in the aliphatic chains of amino acids. The elution order of omega-amino acids and positional isomers of amino pentanoic acid was strongly dependent on the mobile phase pH in the investigated range whereas the retention factors of all alpha-amino acids remained essentially unchanged on all tested stationary phases.