The aim of the study was to perform assessment of radiochromic gel dosimeter based on Turnbull blue dye formed by irradiation (TB gel dosimeter) for measurement of ROFs for 4 mm and 8 mm collimators for the Leksell Gamma Knife PerfexionTM. All measurements have been carried out using home-made spherical Perspex glass phantom of diameter 160 mm TB gel dosimeters were scanned using homemade optical CT scanner.
The results are compared with vendor recommended Monte Carlo calculated ROFs values of 0.814 and 0.900 for 4 mm and 8 mm collimators, respectively. The comparisons between the gel measurements and the treatment planning system (TPS) calculation are presented in the form of 2D isodoses for the central slices and 1D profile.
Measured ROF 0.746 and 0.874 for 4 mm and 8 mm collimators respectively are in a reasonable agreement with vendor recommended values and measured relative dose distribution in a central slice and measured profiles of all shots show excellent correspondence with TPS.