Endothelial Dysfunction Expressed as Endothelial Microparticles in Patients With End-Stage Heart Failure

Publikace na 1. lékařská fakulta |
2014

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

Left ventricular assist devices (LVAD), currently used in treatment of terminal heart failure, are working on principle of rotary pump, which generates continuous blood flow. Non-pulsatile flow is supposed to expose endothelial cells to high stress and potential damage.

Therefore, we investigated longitudinal changes in concentration of circulating endothelial microparticles (EMP) as a possible marker of endothelial damage before and after implantation of LVAD. Study population comprised 30 patients with end-stage heart failure indicated for implantation of the Heart Mate II LVAD.

Concentrations of microparticles were measured as nanomoles per liter relative to phosphatidylserine before and 3 months after implantation. At 3 months after implantation we observed significant decrease in concentration of EMP [5.89 (95 % CI 4.31-8.03) vs. 3.69 (95 % CI 2.70-5.03), p= 0.03] in the whole group; there was no difference observed between patients with ischemic etiology of heart failure (n= 18) and with heart failure of non-ischemic etiology (n = 12).

In addition, heart failure etiology had no effect on the rate of EMP concentration decrease with time. These results indicate possibility that LVAD do not cause vascular damage 3 months after implantation.

Whether these results suggest improvement of vascular wall function and of endothelium is to be proved in long-term studies.