Charles Explorer logo
🇨🇿

Evaluation of new low-cost particle monitors for PM2.5 concentrations measurements

Publikace na Ústřední knihovna |
2017

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

Sixty-six, new, low-cost Speck monitors were collocated with a Grimm 1.109 particle spectrometer and CO loggers during one indoor and two outdoor monitoring campaigns to assess the ability of the monitors to measure the PM concentrations under real use conditions. The limit of detection of the monitors was calculated, and the monitors' bias and precision, as well as the results' reproducibility, were determined.

Correlation coefficients of 0.3 and 0.2 observed between the Speck units and the Grimm instrument during indoor and outdoor measurements were increased by separating the data into sets affected by different sources. Between the Specks, the correlation was >0.9 indoors, showing a high degree of reproducibility among the results.

The limit of detection was calculated using the method of Wallace et al. (2011) and was found to be 10 μg mMINUS SIGN 3 at a 99% confidence level for 1-minute data. The mean bias values showed a PM overestimation of 200% for indoor data, and 500% for outdoor data relative to the Grimm values.

However, these values are influenced by the low measured PM concentrations. The RMSE of Specks was about 10 μg mMINUS SIGN 3 and the concentrations were often about this value.

The overall precision was 8.8+-1.7% indoors and 12.2+-6.2% outdoors, suggesting its independence from changing ambient conditions. The Speck monitors may be suitable for use in indoor and outdoor PM monitoring programs, depending on the required performance to meet the goals of the study.

The instrument has a large bias at low concentrations, making it unsuitable for background measurements or for measurements in clean environments. Even at concentrations over the LOD, it is necessary to assess the bias for expected particle sources before any instrument deployment.